FLIR ThermaCAM EX320 User Manual page 125

Table of Contents

Advertisement

Step
Action
3
Measure the radiation intensity (= apparent temperature) from the reflecting source
using the following settings:
Emissivity: 1.0
D
You can measure the radiation intensity using one of the following two methods:
10589003;a2
Figure 16.3 1 = Reflection source
➲ Please note the following:
Using a thermocouple to measure reflecting temperature is not recommended for
two important reasons:
A thermocouple does not measure radiation intensity
A thermocouple requires a very good thermal contact to the surface, usually by
gluing and covering the sensor by a thermal isolator.
16.2.1.1.2
Method 2: Reflector method
Step
Action
1
Crumble up a large piece of aluminum foil.
2
Uncrumble the aluminum foil and attach it to a piece of cardboard of the same
size.
3
Put the piece of cardboard in front of the object you want to measure. Make sure
that the side with aluminum foil points to the camera.
4
Set the emissivity to 1.0.
Publ. No. 1558146 Rev. a156 – ENGLISH (EN) – February 28, 2006
: 0
obj
16 – Thermographic measurement techniques
16
111