34
User Guide
No.
85*
86
87*
88
95
96
97
100
P/N 300862-K
11/2019
Type
Message
Warning
Device temperature very low
Error
Device temperature too low
Warning
Device temperature very high
Error
Device temperature too high
Warning
A X-gas mixer is connected, must be X-gas
Error
Mixer high pressure alarm on gas input X
Warning
Pressure alarm on mixer X gas input X
Warning
Cannot bridge to a mixer with older software version
EN
Dansensor
MAP Mix Provectus
®
User action
Place unit in
temperature
above 0 °C
Allow unit to
heat up
Place unit in
temperature
above 0 °C
Allow unit to
heat up
Check cooling/
filters
If error persist,
contact service
Check cooling/
filters
If error persist,
contact service
Check settings
Slave mixer must
have same
number of
gasses and types
as Master mixer.
Input pressure
over 10.5 bar
Adjust input
pressure to less
than 10 bar
Check gas supply
Use mixer with
compatible SW
version
Need help?
Do you have a question about the mocon Dansensor MAP Mix ProVectus and is the answer not in the manual?
Questions and answers