RayTek 3i Operator's Manual

RayTek 3i Operator's Manual

Raytek operator's manual digital thermometer 3i
Hide thumbs Also See for 3i:

Advertisement

®
RAYNGER
3i
SERIES
OPERATOR'S MANUAL
Rev H
6/98
56700-1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3i and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for RayTek 3i

  • Page 1 ® RAYNGER ™ SERIES OPERATOR’S MANUAL Rev H 6/98 56700-1...
  • Page 3 In the event of failure of a product covered by this warranty, Raytek will repair the instrument when it is returned to an authorized Service Facility within one year of the original purchase, provided the warrantor’s examination discloses to its satisfaction that the product was defective.
  • Page 5: Table Of Contents

    DATA OUTPUTS ...2-23 2.6.1 Digital Output...2-24 2.6.1.1 Data Output—RUN Loop ...2-24 2.6.1.2 Data Output—SET Loop ...2-25 2.6.1.3 Data Output—LOG Run Loop ...2-26 2.6.1.4 Data Output—LOG SETUP Loop ...2-27 2.6.2 Analog Output...2-28 SECTION Raynger 3i Series Operator's Manual PAGE PAGE Table of Contents...
  • Page 6: Section Page

    OPERATIONAL ...3-2 ELECTRICAL ...3-3 ENVIRONMENTAL ...3-4 PHYSICAL ...3-4 DEFAULT VALUES ...3-5 REGULATORY ...3-6 MAINTENANCE...4-1 BATTERY REPLACEMENT ...4-1 CLEANING ...4-1 4.2.1 Front Window Cleaning ...4-1 4.2.2 Cleaning the Housing...4-2 LASER MAINTENANCE ...4-2 SECTION Table of Contents Raynger 3i Series Operator's Manual PAGE...
  • Page 7 TYPICAL EMISSIVITY VALUES...B-2 APPENDIX C: TROUBLESHOOTING ... APPENDIX D: OPTIONS AND ACCESSORIES... APPENDIX E: TRACEABILITY OF INSTRUMENT CALIBRATION ...E-1 APPENDIX F: CE CONFORMITY FOR THE EUROPEAN COMMUNITY ...F-1 GLOSSARY OF TERMS INDEX Raynger 3i Series Operator's Manual Table of Contents...
  • Page 8 Table of Contents Raynger 3i Series Operator's Manual...
  • Page 9: Introduction

    They can also measure product temperatures during manufacturing or storage without contaminating or marring the product. Table 1-1 lists the standard Raynger 3i models. Table 1-1: Raynger 3i Models TEMPERATURE MODEL RANGE LTDL2 &...
  • Page 10: Inventory

    • Crossed laser—the point where the two laser beams cross is the location of the minimum diameter measurement spot. • Single laser with scope INVENTORY Your Raynger 3i package contains the following: • Raynger 3i • Carrying Case • Operator’s Manual •...
  • Page 11: Model Identification

    XXX is an abbreviation of the company name, YYY is the product (or abbrevia- tion of the product name), and ZZZZZ is the model type. (Model type may be four or five characters long. Refer to Table 1-1 to compare the label to model type.) Raynger 3i Series Operator’s Manual...
  • Page 12 Raynger 3i Series Operator’s Manual...
  • Page 13: Operation

    3. For the short wavelength units (e.g., 2 µm and below)—Avoid taking temperature measurements in bright sunlight. High levels of ambi- ent light may produce apparently valid high-temperature readings when no target is in the thermometer’s field-of-view. Raynger 3i Series Operator’s Manual WARNING—LASER SAFETY IMPORTANT...
  • Page 14 LOG RCL SET °C °F Raynger 3i Series Operator’s Manual...
  • Page 15: Principles Of Operation

    PRINCIPLES OF OPERATION An infrared thermometer and the human eye are very similar. An infrared ther- mometer has a lens that focuses infrared radiation from an object onto a detector. The eye focuses light onto the retina. The detector is stimulated by the incoming infrared energy and produces a signal that is transmitted to the circuitry.
  • Page 16: Your Portable Infrared Thermometer

    Celsius or Fahrenheit. A digital/analog output allows data recording, instrumentation or process control, and/or remote display of temperature and emissivity. The instru-6(. The instr)7.6con Raynger 3i Series Operator’s Manual...
  • Page 17 • Analog Output—Connects the instrument to analog recording/printing devices such as chart recorders and printers. • Digital Output—An RS232 interface to connect the instrument to a computer or directly to a printer’s RS232 port. • DC In—AC adaptor connection. Raynger 3i Series Operator’s Manual...
  • Page 18: Operation And Controls

    Remove the batteries if the unit is not used for long periods of time. 220 V AC adaptors must have DIN VDE 0551 approval to be used with IEC Class 2 laser units. HOURS OF CONTINUOUS USE 12.5 12.5 1 year Raynger 3i Series Operator’s Manual...
  • Page 19: Control Panel And Display

    Laser on/off button Data logger icon Trigger lock icon Backlight icon Mode display Laser on/off LED Figure 2-4: The Control Panel and Display Raynger 3i Series Operator’s Manual LOG RCL SET C F RECALL MODE ACTIVATE ACTIVATE • • • •...
  • Page 20 4.3 V, the laser will automatically turn off. Also, if the unit is in the LOCK mode, the laser will go off when the trigger is released. Temperature display—Shows the current temperature (while the trigger is pulled) or the last temperature measured (when the RECALL button is pressed). Raynger 3i Series Operator’s Manual...
  • Page 21 Trigger lock icon—The small padlock is the trigger lock icon and is activated when LOCK is pressed while the trigger is pulled. To unlock the trigger, simply press the LOCK button again. Raynger 3i Series Operator’s Manual...
  • Page 22: Control System

    LASER °C °F HI ( ) LO ( ) T amb MODE ACTIVATE • • RECALL LOCK LASER °C °F HI ( ) LO ( ) T amb MODE ACTIVATE • • RECALL LOCK LASER Raynger 3i Series Operator’s Manual...
  • Page 23: Control Loops

    For the short wavelength units (e.g., 2 µm and below)—Avoid taking temperature measurements in bright sunlight. High levels of ambient light may produce apparently valid high-temperature readings when no target is in the thermometer’s field-of-view. Raynger 3i Series Operator’s Manual IMPORTANT 2-11...
  • Page 24: Run Loop-To Measure Temperature

    °F MODE Change emissivity with the • buttons in any mode BUTTON LOOP ICON Figure 2-6: RUN Loop Raynger 3i Series Operator’s Manual MIN Mode °C °F HI ( ) LO ( ) T amb RECALL MODE ACTIVATE ACTIVATE •...
  • Page 25 RUN loop, the instrument is reset to its RUN loop factory default settings (stored data is unaffected). The instrument will “beep” after reinstalling the default settings. Section 3.6 lists default setting for each model as well as non-model-specific default set- tings and range values. Raynger 3i Series Operator’s Manual + ...R = AVG 2-13...
  • Page 26: Log Loop-To Measure And Store Temperature

    RECALL MODE the log location (LOC) number BUTTON LOOP ICON Figure 2-7: LOG Loop buttons . Raynger 3i Series Operator’s Manual MAX Mode °C °F HI ( ) LO ( ) T amb ACTIVATE MODE ACTIVATE •...
  • Page 27 LOG loop factory default settings. The instrument will “beep” after reinstalling the default settings. Section 3.6 lists default setting for each model as well as non-model-specific default settings and range values. Raynger 3i Series Operator’s Manual IMPORTANT 2-15...
  • Page 28: Recall Loop-To Recall Measured Temperature

    Recall button again to continue the Recall function °C °F HI ( ) LO ( ) T amb ACTIVATE • LOCK LASER MODE °C °F HI ( ) LO ( ) T amb ACTIVATE • LOCK LASER Raynger 3i Series Operator’s Manual...
  • Page 29: Recall Values From Log

    Press the Recall button again, if display goes out, to continue Recall function TRIGGER Released & Unlocked Figure 2-9: RECALL Loop–Recalling LOG Values Raynger 3i Series Operator’s Manual MODE Stored temperature. If no temperature is stored, the display shows all zeros.
  • Page 30 RECALL contains two modes: LOC (location) and MAX (as shown in Figure 2-9). In RECALL, the stored temperature and emissivity setting along with either the location number or MAX temperature can be displayed for each of the 100 LOG locations. 2-18 Raynger 3i Series Operator’s Manual...
  • Page 31: Setup Loop-To Setup And Activate Alarms And Features

    LASER Press the ACTIVATE LASER button to toggle between DIG or ANA DIG/ANA Mode TRIGGER Released & Unlocked Figure 2-10: SETUP Loop–RUN Values Raynger 3i Series Operator’s Manual °C °F MODE ACTIVATE RECALL Press the ACTIVATE • • button to toggle...
  • Page 32 Step 7 below. Release the trigger. 5. Press the SET button (do not pull the trigger) to enter the Setup mode 6. Press the MODE button until TAM shows in the mode function indicator. 2-20 Raynger 3i Series Operator’s Manual...
  • Page 33: Setup Values For Log

    2. Press the SET button. The SET icon will be activated. 3. Press the RUN/LOG button, if necessary, so that the LOG icon is activated. 4. Press the buttons to select the LOG location number (LOC mode). Raynger 3i Series Operator’s Manual buttons. HAL Mode MODE...
  • Page 34: Using A Tripod

    LOCK button. Note that battery life is shortened by continuous use. If you plan to monitor tempera- tures over long intervals, you should connect an AC adaptor to the DC IN connector. 2-22 Raynger 3i Series Operator’s Manual...
  • Page 35: Data Outputs

    Baud Rate: 9600 Data Format: 8 bits, 1 stop bit, no parity • Analog: 1 mV/° (°C or °F) for all models except the 1M 1 mV/°C or 0.5 mV/°F for the 1M Raynger 3i Series Operator’s Manual Analog RS232 RS232...
  • Page 36: Digital Output

    0090 0090 0070 0020 088.4 –11.5 100.0 999.9 1000 0090 –018 –018 –028 –028 Raynger 3i Series Operator’s Manual Notes Carriage Line (see below) Return Feed <CR> <LF> <CR> <LF> <CR> <LF> <CR> <LF> <CR> <LF> <CR> <LF> <CR> <LF>...
  • Page 37: Data Output-Set Loop

    °F, and DOI = 3600 seconds. (The <CR> and <LF> do not print; these denote the carriage return and line feed.) Note: DOI does not apply to the SETUP mode. Raynger 3i Series Operator’s Manual (hal, lal, tam not activated) (HAL activated)
  • Page 38: Data Output-Log Run Loop

    DOI does not apply to the LOG mode. 2-26 Stored MAX Location Mode Temperature Number 0090 0100 0090 HAL <CR> 0100 Raynger 3i Series Operator’s Manual Notes Carriage Line (see below) Return Feed <CR> <LF> <CR> <LF> <LF> <CR>...
  • Page 39: Data Output-Log Setup Loop

    3. For Locations 3, 4, 5, 6, and 7, HAL is not activated and is 85°F. 4. For Location 8, HAL is activated and is 65°F. Note: DOI does not apply to the LOG SETUP mode. Raynger 3i Series Operator’s Manual °C Location...
  • Page 40: Analog Output

    Use the signal output jack accessory. The analog output represents the current temperature of the object being measured, regardless of the mode used. If T will be representative of the compensated temperature values. 2-28 compensation is activated, the analog output Raynger 3i Series Operator’s Manual...
  • Page 41: Sighting Systems

    Note: The laser turns off automatically when the trigger is released. It cannot be locked on when the trigger is locked. It cannot be turned on in the Recall and Setup loops. Raynger 3i Series Operator’s Manual FDA Class IIIa...
  • Page 42 NE PAS EXPOSER L'OEIL <1mW/630–670nm AU RAYON LASER IEC 825/93 LASER DE CLASSE 2 Figure 2-13: Laser Labels Raynger 3i Series Operator’s Manual COMPLIES WITH LASER RADIATION - DO FDA 21 CFR NOT STARE INTO BEAM SUBCHAPTER J OUTPUT < 1mW...
  • Page 43: Single Laser Sighting

    DISTANCE: SENSOR TO OBJECT (m) IR Spot Diameter at Lens = 23 mm (0.9 in) Laser Diameter at Lens = 40 mm (1.6 in) Figure 2-15: Dual Laser and IR Spot Diameters Raynger 3i Series Operator’s Manual LASER Far Field D:S = 68:1 2-31...
  • Page 44: Crossed Laser Sighting

    Figure 2-17: Scope Sighting 2-32 Focus Point D:S = 75:1 Far Field D:S = 20:1 DISTANCE: SENSOR TO OBJECT (in) 1500 3000 DISTANCE: SENSOR TO OBJECT (mm) Raynger 3i Series Operator’s Manual 5000...
  • Page 45 The scope has one reticle and it indicates the target area that is measured at the focus point (see Appendix A for the focus point distance of your model). See section 2.7.4 and Figure 2-17 for further details on the scope. Figure 2-18: Scope with Laser Sighting Raynger 3i Series Operator’s Manual 2-33...
  • Page 46 2-34 Raynger 3i Series Operator’s Manual...
  • Page 47: Specifications

    Note: Hi Alarm, Lo Alarm, and T. Ambient ranges are the same as measurement ranges in Table 3-1. Raynger 3i Series Operator’s Manual 600 to 3000 C 200 to 1800 C (1100 to 5400 F) (400 to 3275 F) 1.0 micron...
  • Page 48: Operational

    700 msec (95% response) C or F (selectable), multifunctional 4-digit backlit LCD Membrane Switch Panel 1 to 9999 seconds 1 C or 1 F 1 to 100 550 msec 550 msec Raynger 3i Series Operator’s Manual 700 msec 700 msec...
  • Page 49: Electrical

    Range Power Four (4) AA alkaline or rechargeable batteries or 6 to 9 volt, 200 mA, DC power supply Requirements Raynger 3i Series Operator’s Manual 1mV/ C (0.5mV/ F) Same as Measurement Range in Table 3-1 1.3 K RS232, 9600 baud...
  • Page 50: Environmental

    244 mm (9.6 in) H 257 mm (10.1 in) L 257 mm (10.1 in) L 71 mm (2.8 in) W 1000 g (2.21 lbs) Raynger 3i Series Operator’s Manual 244 mm (9.6 in) H 71 mm (2.8 in) W 1000 g (2.21 lbs)
  • Page 51: Default Values

    Setup Screen (LOG Loop) LOG Location LOG Data–All Locations Current Temperature MAX Temperature Hi Alarm Value Hi Alarm State Emissivity Raynger 3i Series Operator’s Manual Hi Alarm Lo Alarm 1200 C (2200 F) (0 F) 3000 C 600 C (5400 F)
  • Page 52: Regulatory

    The appropriate regulatory approvals and certificates have been issued as follows: • FDA Class II Laser Certification • FDA Class IIIa Laser Certification • IEC Class 2 Laser Certification • IEC801-3 (EN50082) for EMI susceptibility Raynger 3i Series Operator’s Manual...
  • Page 53: Maintenance

    Do not wipe the surface dry, as this may scratch the surface. Do not use any ammonia, or cleaners with ammonia, bleach, acids, or strong bases. This can cause severe damage to the front window. Raynger 3i Series Operator’s Manual CAUTION...
  • Page 54: Cleaning The Housing

    Wipe with a damp sponge or soft rag. Use a soft rag to gently wipe the dis- play. LASER MAINTENANCE If the laser (laser models only) does not operate properly, call your infrared ther- mometer supplier. DO NOT open the instrument’s main housing. Raynger 3i Series Operator’s Manual...
  • Page 55: Appendix A: Optical

    Focus Point D:S = Distance to spot divided by spot diameter at the focus point Far Field D:S = Ratio at distances greater than 10x the focus distance Figure A-1: How to Read the Optical Charts Raynger 3i Series Operator’s Manual 1500 3000...
  • Page 56: Optical Charts

    Laser Diameter at Lens = 40 mm (1.6 in) FAR FIELD D:S = 68:1 IR Beam Profile Laser Profile G5SC Raynger 3i Series Operator’s Manual DISTANCE: SENSOR TO OBJECT IN INCHES Focus Point 0.3 IN @ 24 IN 8 mm @ 610 mm...
  • Page 57 DISTANCE: SENSOR TO OBJECT IN METERS FOCUS POINT D:S = 90:1 FAR FIELD D:S = 60:1 IR Spot Diameter at Lens = 23 mm (0.9 in) Raynger 3i Series Operator’s Manual LRSCL2 FOCUS POINT D:S = 105:1 @ 15 m (50 ft) 3000...
  • Page 58 Raynger 3i Series Operator’s Manual...
  • Page 59: Appendix B: Object Emissivity

    0.93. Finally, measure an adjacent area on the object and adjust the emissivity setting until you reach the same temperature. This is the correct emissivity for the measured material. Raynger 3i Series Operator’s Manual...
  • Page 60: Typical Emissivity Values

    0.03 n.r. 0.05-0.2 0.2-0.8 0.2-0.9 n.r. n.r. 0.01-0.1 0.5-0.9 0.6-0.9 0.4-0.9 0.6-0.9 0.3-0.4 0.3-0.6 0.2-0.5 0.25 Raynger 3i Series Operator’s Manual 8-14µm n.r. 0.2-0.4 0.1-0.3 n.r. n.r. n.r. n.r. n.r. 0.4-0.8 n.r. 0.3-0.8 0.7-0.95 0.3-0.6 0.15 Continued on next page...
  • Page 61 Polished Sheet Molten Oxidized Stainless Tin (Unoxidized) Titanium Polished Oxidized Tungsten Polished Zinc Oxidized Polished Material n.r. = not recommended Raynger 3i Series Operator’s Manual Emissivity 1.0µm 1.6µm 0.4-0.8 0.5-0.9 0.35 0.1-0.3 n.r. 0.6-0.9 0.35 0.4-0.6 0.7-0.9 0.7-0.9 0.35 0.35 0.3-0.4...
  • Page 62 — 0.98 0.4-0.98 0.98 — 0.9-0.95 0.95 0.95 0.95 0.95 0.95 — — 0.9-0.98 — 0.93 0.9-0.95 0.9-0.95 Raynger 3i Series Operator’s Manual 8-14µm 0.95 0.95 0.8-0.9 0.7-0.8 0.95 0.95 0.95 0.95 0.85 n.r. 0.95 0.8-0.95 0.98 0.98 0.9-0.95 0.95 0.95...
  • Page 63: Appendix C: Troubleshooting

    RS-232 output are sent the -U- failsafe code. The mode display and RS- 232 ouput are sent either the correct MAX temperature or the -U- failsafe code. The analog output is sent the minimum temperature for your model. Raynger 3i Series Operator’s Manual LOG RCL SET °C °F MODE ACTIVATE •...
  • Page 64 Note: On laser models, if the internal temperature is less than 0°C (32°F) or greater than 45°C (113°F), the laser will automatically turn off. The laser will also automatically turn off if the battery is too low (when the low battery icon displays). Raynger 3i Series Operator’s Manual...
  • Page 65 To clear the instrument’s system, release then press the trigger. If this does not clear the problem, slide the battery door off then back on. (This disconnects the battery power.) Raynger 3i Series Operator’s Manual LOG RCL SET °C °F RECALL...
  • Page 66 Raynger 3i Series Operator’s Manual...
  • Page 67: Appendix D: Options And Accessories

    • Laser glasses OPTIONS A full range of options for various applications and environments is available. Options are factory installed and must be ordered with base model units. The follow- ing options are available: • NIST certification Raynger 3i Series Operator’s Manual...
  • Page 68 Raynger 3i Series Operator’s Manual...
  • Page 69: Appendix E: Traceability Of Instrument Calibration

    1200 - 3000 C (2200 - 5400 F) Certified Radiation Transfer Standard Calibration Source (Blackbody) Emissivity Figure E-1: Traceability Flowchart Raynger 3i Series Operator’s Manual (NIST) standards traceable to NIST Medium Temperature Models 200 - 1200 C (400 - 2200 F) Certified RTD Instrumentation & NIST...
  • Page 70 Raynger 3i Series Operator’s Manual...
  • Page 71: Appendix F: Ce Conformity For The European Community

    The instrument’s average error in this frequency range for the three orientations is 3.1° C at an electric field strength of 3 V/m. At some frequencies the instrument may not meet its stated accuracy. Raynger 3i Series Operator’s Manual...
  • Page 72 Raynger 3i Series Operator’s Manual...
  • Page 73: Glossary Of Terms

    Average. Units with this feature have the abil- ity to display the average of measured temper- atures. AVG is a weighted average of all read- ings taken since the trigger was pulled. Raynger 3i Series Operator's Manual Glossary Of Terms Background Temperature Temperature behind and surrounding the target, as viewed from the instrument.
  • Page 74 Units in which frequency is expressed. Synonymous with cycles per second. International Electrotechnical Commission; a European organization that coordinates and sets standards among the European Community. IEEE-488 A communications format standard. See dig- ital data bus. Raynger 3i Series Operator's Manual...
  • Page 75 Single or dual lasers are used in some units for aiming and/or locating the optimum tem- perature measurement point. Raynger 3i Series Operator's Manual Location. Units with the datalogging feature store data in numbered locations, which can be recalled and reviewed on the display when necessary.
  • Page 76 95 percent of full scale temperature indication (per the ASTM standard test method E 1256-88). The specification for Raytek instruments also includes the average time required for soft- ware computations. Raynger 3i Series Operator's Manual...
  • Page 77 Spectral Filter An optical element used to restrict the spec- tral band of energy reaching an instrument’s detector. Raynger 3i Series Operator's Manual Spectral Response The wavelength region in which the IR Thermometer is sensitive. Spot The diameter of the area on the target where the temperature determination is made.
  • Page 78 An oscillatory or repetitive motion test of the instrument per MIL-STD-810D or IEC 68-2-6, often specified as an acceleration in g’s (1g = 9.81m/s 2 = 32.2 ft./s 2 ), over a frequency range typically measured in Hertz (sec -1 ). Raynger 3i Series Operator's Manual...