JUMO 202710/30 Operating Manual page 21

Handheld meter for conductivity (cr), specific electrical resistivity, tds, and salinity
Hide thumbs Also See for 202710/30:
Table of Contents

Advertisement

(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
NOTE!
In the following list of parameters, the setting options for the default setting
are shown in bold in each case.
Parameter "InP": selection of a measurand
Select the desired measurand with the
Menu/parameter
Press the
press the
setting menu or
press the
measurement mode.
All manuals and user guides at all-guides.com
Parameters
"InP": selection of a measurand, see page 21
"C.tdS": conversion factor for TDS measurement, see page 22
"CELL Corr": setting the relative cell constant, see page 22
"rAnG": selection of the display range or the measuring range, see page 23
"CAL": preconfiguration for the calibration of the relative cell constant,
see page 23
"rEF.S": selection of the standard reference solution for the calibration of the
relative cell constant, see page 24
"Unit t": temperature unit °C/°F, see page 24
"t.Cor": setting the temperature compensation, see page 25
"t.Lin": compensation coefficient, see page 25
"t.rEF": setting the reference temperature, see page 26
"C.int": calibration timer, see page 26
"Auto HLD": automatic saving of a measured value, see page 27
"P.oFF": automatic device switch-off, see page 27
Setting option
Cond
rESi
tdS
SAL
SEt
key to save the setting and move to the next parameter or
key to save the setting and move to the next submenu of the
key to save the setting, quit the device settings, and return to
6 Device settings
and
keys:
Meaning
Conductivity
Specific electrical resistivity
Filtrate dry residue
Salinity (salt content)
Selection of the measurand in measure-
ment mode using the "Set" key
21