8.4

Parameterizing the device

8.4.1 Submenu "SEt Corr": offset/zero point correction and slope cor-
rection of the voltage and temperature measurement
Overview of the parameters
Parameter "OFFS": offset/zero point correction of the voltage measurement
After calling up the setting menu with
"SEt Corr", use
to select the parameter to be configured.
rH
mV
mV
pH
H
ATC
ok
CAL
AL Logg
pH
Parameters
(3)
"OFFS mV": offset/zero point correction of the voltage measurement, See
page 41
(4)
"SCAL mV": slope correction of the voltage measurement, See page 42
(3)
"OFFS °C": offset/zero point correction of the temperature measurement, See
page 42
(4)
"SCAL °C": slope correction of the temperature measurement, See page 43
Use the
and
Menu/parameter
rH
mV
mV
pH
H
ATC
ok
CAL
AL Logg
pH
Press the
key to save the setting and move to the next parameter or
press the
key to save the setting and move to the next submenu of the
setting menu or
press the
key to save the setting, quit the device settings, and return to
measurement mode.
(1)
mV
CAL
3
(2)
mV
keys to set the zero point correction:
Setting option
Meaning
oFF
no zero point correction of the voltage
measurement
-
zero point correction of the
10.0 to +10.0 m
voltage measurement in mV
V