Wireless communication interference
This product operates in an unlicensed ISM band at 2.4 GHz. In the event this product
is used near other wireless devices such as microwave and wireless LAN, which
operate on the same frequency band as this product, there is a possibility that
interference may occur. If interference occurs, stop the operation of the other devices
or relocate this product away from other wireless devices before attempting to use it.

8. PC-Link Functions

This device can be used in conjunction with a personal computer (PC) running the
Microlife Blood Pressure Analyser (BPA+) software. The memory data can be
transferred to the PC by connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA software from
www.microlife.com/technologies/connect/bpaplus

9. Error Messages

If an error occurs during the measurement, the measurement is interrupted and an
error message, e.g. «ERR 3», is displayed
Error
Description
«ERR 1»
Signal too
21
weak
«ERR 2»
BP Error signal During the measurement, error signals were
20-B
«ERR 3»
BM No
20-C
pressure
in the cuff
«ERR 5»
Abnormal
result
«ERR 6»
MAM Mode
Potential cause and remedy
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the measurement.*
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
An adequate pressure cannot be generated in the
cuff. A leak may have occurred. Check that the cuff
is correctly connected and is not too loose. Replace
the batteries if necessary. Repeat the measurement.
The measuring signals are inaccurate and no result
can therefore be displayed. Read through the
checklist for performing reliable measurements and
then repeat the measurement.*
There were too many errors during the
measurement in MAM mode, making it impossible to
obtain a final result. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*