Microlife BP W10 Manual
Microlife BP W10 Manual

Microlife BP W10 Manual

Wrist blood pressure monitor

Advertisement

Quick Links

Microlife AG
Microlife UAB
Espenstrasse 139
P. Lukšio g. 32,
9443 Widnau / Switzerland
08222 Vilnius
www.microlife.com
Lithuania
IB BP W10 EN Master 2022
Revision Date: 2022-04-26
BP
W10
Wrist Blood Pressure Monitor
EN
1

Advertisement

Table of Contents
loading

Summary of Contents for Microlife BP W10

  • Page 1 Wrist Blood Pressure Monitor Microlife AG Microlife UAB Espenstrasse 139 P. Lukšio g. 32, 9443 Widnau / Switzerland 08222 Vilnius www.microlife.com Lithuania IB BP W10 EN Master 2022 Revision Date: 2022-04-26...
  • Page 2 Microlife BP W10 Guarantee Card BP W10 Name of Purchaser Serial Number Date of Purchase Specialist Dealer...
  • Page 3 Microlife BP W10 AP Stored value 1 ON/OFF button AQ Battery display 2 Display AR Cuff fit check 3 M-button (memory) -A: Suboptimal cuff fit 4 Time button -B: Arm movement indicator «ERR 2» -C: Cuff pressure check «ERR 3»...
  • Page 4 (e.g. home-use). Humidity limitation Dear Customer, Temperature limitation Your new Microlife blood pressure monitor is a reliable medical device for taking measurements on the wrist. This device was devel- oped in collaboration with physicians and clinical tests prove its Medical device...
  • Page 5: Table Of Contents

    6. Data memory please contact your local Microlife-Customer Service. Your dealer or Viewing the stored values pharmacy will be able to give you the address of the Microlife dealer Clearing all values in your country. Alternatively, visit the internet at www.microlife.com 7.
  • Page 6: Important Facts About Blood Pressure

    1. Important facts about blood pressure in the evening: before going to bed, bathing or taking medication) and average the measurements.  Blood pressure is the pressure of the blood flowing in the arteries  It is quite normal for two measurements taken in quick succession generated by the pumping of the heart.
  • Page 7: Using The Device For The First Time Inserting The Batteries

    13.5 - 21.5 cm (5.25 - 8.5 inches). button for approx. 3 seconds in the setting mode. Press time button again and then the year number starts to flash. Now you can enter the new values as described above. Microlife BP W10...
  • Page 8: Taking A Blood Pressure Measurement

    4. Taking a blood pressure measurement  You can stop the measurement at any time by pressing the ON/OFF button (e.g. if you feel uneasy or an unpleasant pres- 1. Press the ON/OFF button 1 to start the measurement. sure sensation). 2.
  • Page 9: Appearance Of The Irregular Heartbeat (Ihb) Symbol

    «CL» appears and then release the button. To cates when the heart rate is irregular. permanently clear the memory, press the M-button while «CL» is flashing. Individual values cannot be cleared. Microlife BP W10...
  • Page 10: Battery Indicator And Battery Change

     Cancel deletion: press ON/OFF button 1 while «CL» is Which batteries and which procedure? flashing.  Use 2 new, long-life 1.5 V, size AAA alkaline batteries.  Do not use batteries beyond their date of expiry. 7. Battery indicator and battery change ...
  • Page 11: Error Messages

     This device comprises sensitive components and must be treated able measurements and then repeat the with caution. Observe the storage and operating conditions measurement.* described in the «Technical specifications» section. Microlife BP W10...
  • Page 12  The cuffs are sensitive and must be handled with care.  Presence of significant cardiac arrhythmia during measurement  Only pump up the cuff once fitted. may interfere with blood pressure measurement and affect the  Do not use this device if you think it is damaged or notice anything reliability of blood pressure readings.
  • Page 13  Prolonged exposure to cuff pressure will reduce peripheral perfu- sion and may lead to injury. Avoid situations of extended cuff pres- Microlife BP W10...
  • Page 14: Device Care

     The device is intended only for measuring blood pressure at wrist. - direct sunlight Do not measure other sites because the reading does not reflect - contamination and dust your blood pressure accurately.  This device has 2 user settings. Please ensure it is cleaned and ...
  • Page 15: Accuracy Test

    (e.g. being dropped). Please contact your Should guarantee service be required, please contact the dealer local Microlife-Service to arrange the test (see foreword). from where the product was purchased, or your local Microlife Disposal service. You may contact your local Microlife service through our...
  • Page 16: Technical Specifications

    11.Technical specifications Reference to EN 1060-1 /-3 /-4; IEC 60601-1; standards: IEC 60601-1-2 (EMC); IEC 60601-1-11 Expected service Device: 2 years or 10000 measurements, Operating 10 - 40 °C / 50 - 104 °F life: whichever comes first conditions: 15-90 % relative maximum humidity Accessories: 2 years or 10000 measure- Storage conditions: -20 - +55 °C / -4 - +131 °F ments, whichever comes first...

Table of Contents