However, if the symbol appears on a regular basis (e.g., several times a week with
measurements taken daily), we advise you to tell your doctor.
Please show your doctor the following explanation:
Information for the doctor on frequent appearance of the irregular heartbeat
symbol
This instrument is an oscillometric blood pressure monitor device that also analyzes
pulse frequency during measurement. The instrument is clinically tested. If pulse
irregularities occur during measurement, the irregular heartbeat symbol is displayed
after the measurement. If the symbol appears more frequently (e.g., several times per
week on measurements performed daily) or if it suddenly appears more often than
usual, we recommend the patient to seek medical advice.
The instrument does not replace a cardiac examination but serves to detect pulse
irregularities at an early stage.
4. Data Memory
This device automatically stores up to 99 measurement values for each of the 2 users.
Select either user1 or 2 by pressing the user button 5
Viewing the average of the last 28 days
Press the Memory button 3 briefly, when the device is switched off. The display first
shows «M» 25 and previous «28A», which stands for the average of measurement
values of the last 28 days.
* Blood pressure readings with suboptimal cuff fit 20-A are not considered in the
average value.
Viewing the clinical blood pressure average « MyBP »
Press the M-button again, allows you to see the clinically relevant blood pressure
average« MyBP ». The display first shows «M» 25 and «MyBP» 26. This clinical blood
pressure average is only shown when 12 clinically relevant measurement values in
the last 28 days have been performed.
* Only measurement that were performed in the morning between (05:00-10:59) or in
the evening (17:00-22:59)
* A maximum of 4 readings per day are considered (2 from the morning and 2 from the
evening)
* Measurements performed in standard mode and in MAM-mode are considered in
the average, when they have been performed during the right time of the day.
* Measurements performed in MAM-modes or single standard mode are both classed
as single measurement for working out the «MyBP average».
* Blood pressure reading with suboptimal cuff fit 20-A are not considered in the
average value.