4.CALIBRATION
4. CALIBRATION
Calibration is performed for matching the F600AT to a loadcell. The following two types of calibration
are available for the F600AT.
Actual load calibration
Apply an actual load to the loadcell and enter the actual load value by the keys for calibration.
Calibration is accurately performed with reductions in errors.
Actual Load
Loadcell
Equivalent input calibration
Calibration is performed without an actual load by entering the rated output value (mV/V) and the
capacity (to be indicated) of the strain-gage sensor by the keys. Calibration is easily performed when
no actual load is available.
For example, the gain is automatically determined by entering:
as indicated for a load.
Loadcell
A data sheet is attached to a loadcell at the time of purchase.
The data sheet provides data including:
Capacity
Rated Output
non-linearity, hysteresis,input resistance,output resistance and zero balance.
Enter the capacity and the rated output value required for equivalent input
calibration into the F600AT.
16
2.001mV/V - 100.0kg
In case of the load
Rated Output
Rated output value (mV/V) + Indicated value
F600AT
Indicated value
capacity
F600AT
Load (in kg, t, etc.)
voltage (in mV/V)