D-ONE – Operating instructions V01
3.5.2
Calibration & Service
These options enable you to set calibration features, review service history and check the
number of pipetting moves performed.
Calibration
Description
& Service
Calibration
Allows for re-calibration of the D-ONE module to restore
accuracy (calibration has to be performed by INTEGRA).
Choose the channel. The calibration factors for Pipette
and Repeat type are displayed.
To edit the calibration volumes, press Edit.
• Target Volume: This is the volume you are interested
• Actual Volume: This is the measured volume
• Current Factor: Displays the factor currently in use.
• Factory Reset: Resets the correction factor back to
Calibration
Sets a calibration reminder based on a specified time
Reminder
frame or number of pipetting cycles. When the calibration
reminder is displayed, press any key to confirm.
However, the reminder will reappear every time the D-
ONE module is turned on until you change the reminder
time or use the reset option.
Time or
• Reminder: Press OK to turn the reminder timer On or
Cycles
• Days/Cycles: Use the touch wheel to set a reminder
• Remind in/Total Cycles: Displays the residual time
• Reset: Resets the timer to the defined calibration
Press to save.
Service
Displays notes of any service that took place on the D-
History
ONE module listed newest entry first.
18
in using for the calibration.
obtained when dispensing the target volume.
the original factory setting. Press to apply the
factory setting.
Off.
interval for calibration (time in days or in thousands of
cycles). Press to set the timer to the defined cali-
bration interval.
or amount of cycles respectively before calibration is
required.
interval. Press to enable.
Installation
Range
-
/
(On/Off)
1-365 days
or 1k - 240k
cycles
-
INTEGRA Biosciences AG