If Output; Video Output; Aux. Port; Aux. Control - Rohde & Schwarz FSVA3000 User Manual

Signal and spectrum analyzer
Hide thumbs Also See for FSVA3000:
Table of Contents

Advertisement

®
R&S
FSVA3000/ R&S
Connector
REF Input 1
REF Output 1
REF Input 2*)
REF Output 2 *)
*) requires option R&S FSV3-K703
3.3.2.9

IF output

The female BNC connector can be used to output the intermediate frequency (IF).
3.3.2.10

Video output

The female BNC connector can be used to provide video output (1 V).
3.3.2.11

Aux. port

A 9-pole SUB-D male connector used to provide low-voltage TTL control signals (max.
5 V). The output signals can be used to control external devices.
3.3.2.12

Aux. control

A 9-pole SUB-D female connector used as input for low-voltage TTL control signals
(max. 5 V) from a controlling external device, such as an external generator.
This connector is provided by the "Additional Interfaces" option R&S FSV3-B5.
User Manual 1178.8520.02 ─ 08
®
FSV3000
Reference signal
1 MHz - 100 MHz (in
1 ppm steps)
0 dBm - 15 dBm
1 MHz - 100 MHz
> 0 dBm
10 MHz, 100 MHz,
128 MHz, 640 MHz,
1000 MHz,
1280 MHz
3 dBm - 13 dBm
640 MHz
10 dBm
Usage
To provide an external reference signal to the R&S FSV/A.
To provide the internal reference signal from the R&S FSV/A
to another device continuously.
Also used to provide OCXO reference signal to another
device.
To provide an external reference signal to the R&S FSV/A.
To provide a 640 MHz reference signal from the R&S FSV/A
to another device.
Getting started
Instrument tour
69

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsv3000

Table of Contents