7.1.3 Transmitting ECG Data in SCP/FDA-XML/PDF/DICOM Format
Through Ethernet Cable
NOTE: The SCP/FDA-XML/DICOM function can be activated on the Advanced Setup
screen. For details, please contact the manufacturer or the local distributor.
1. Log into the FTP receiving software.
2. Connect the electrocardiograph to the PC with an Ethernet cable recommended by the
manufacturer. If WIFI has been used, connect the PC and electrocardiograph to WIFI.
3. Configure the Transmission Setup window.
1) Press Shift+F1 in the Transmission Setup window to open the Basic Setup window.
2) Set Auto Transmission to On and Transmission Mode to Wired or Wireless.
3) Set the Server IP item to the IP of the PC.
For details, please refer to Section 10.6.1 "Basic Setup".
4) Set the FTP User Name, FTP Password and FTP Path items.
a) The user name and the password you input in the FTP User Name and FTP
Password items must be available for FTP server.
b) The path you input in the FTP Path item must be the subdirectory of the path you
input in the FTP receiving software.
NOTE: For more information about FTP server, consult your Network
Administrator.
3. Set File Format to DICOM/SCP/FDA-XML/PDF in the File Setup window.
4. In the auto or rhythm mode, ECG data will be transmitted through the network automatically
after an ECG report is printed out.

7.2 Real-time Transmission to SE-1515

NOTE: To use the real-time transmission function, the SE-1515 software of the
manufacturer must be installed in the PC.
1. Double-click on the SE-1515 shortcut icon to start up the SE-1515 software..
1) Enter the resting ECG sampling screen and click Setup. The System Setting window
will be display.
2) Select the Sampling Setting tab, and set Device Model to SE12 in the Sampling Box
setup panel.
- 67 -