SE-1200 Series Electrocardiograph User Manual
Operation Preparations
Disposable Electrode:
Alligator Clip:
Disposable electrodes must be used together with alligator clips.
Disposable Electrode Connection
1) Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips to
the lead wires.
2) Clean the electrode areas on the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Clip the disposable electrodes with the alligator clips.
The quality of ECG waveform will be affected by the contact resistance between the patient and
the electrode. In order to get a high-quality ECG, the skin-electrode resistance must be minimized
while connecting electrodes.

3.4 Inspection Before Power-On

In order to avoid safety hazards and get good ECG records, the following inspection procedures
are recommended before operation.
WARNING
The electrocardiograph is intended to be used by qualified physicians or personnel
professionally trained, and they should be familiar with the contents of this user manual
before operation.
- 28 -