Ceiling Mount Installation; (A) Drop Ceiling Applications - Labomed Prima DNT User Manual

Operating surgical microscopy
Hide thumbs Also See for Prima DNT:
Table of Contents

Advertisement

5. CEILING MOUNT INSTALLATION

5A. DROP CEILING APPLICATIONS
Many newer constructions use drop ceilings. Whenever this type of application is used, all supporting
structure and electrical service (115/230 VAC, 15A ) is the customer's responsibility and must be completed
before the installation.
WARNING: THE SUPPORTING STRUCTURE FOR THE CEILING MOUNT AND WALL MOUNT INSTALLATION
MUST BE CAPABLE OF SUPPORTING A MINIMUM WEIGHT OF 600 LBS (272 kg) AND BE STABLE
LATERALLY AND VERTICALLY, AND FREE FROM VIBRATION.
WARNING:
• CEILING MOUNT PLATE AND COLUMN ASSEMBLY TOGETHER CAN WEIGHTS APPROXIMATELY UPTO
66LBS (30KG) WITH MICROSCOPE.
• TO AVOID PERSONAL INJURY TWO (2) PEOPLE ARE REQUIRED FOR PROPER INSTALLATION.
PROCEDURE:
1. Ensure that there is a 350mm diameter hole in the drop ceiling to
mount the master Ceiling plate.
2. Labomed Ceiling Mount system for wooden ceiling can be installed
by fixing a flat wooden membrane at least 90mm thick to the joist
and then mount the Master Ceiling Plate to the wooden membrane
as shown in Fig.10 .
3.Using the Paper template provided, mark the positions of the Eight (8)
mounting holes on wooden membrane refering figure.
4. Drill Eight (8) 1/4"(6mm) pilot holes either direct into Joist or the
additional wooden membrane as per the roof available.
5. Place a 10mm flat washer over each of the Eight M10 X 100mm hex
head lag screws.
6. Referring Flat side of Master Plate, line up the (6) holes with the pilot
holes in the mounting surface Zand insert the hex head lag screws into the
holes.
7. Tighten the master Plate with a 17mm Hex Socket Spanner.
8. Position and align the ceiling Mount against the master plate and insert
6 socket head bolts M12 X 40.
9. SlightlyTighten all the bolts using Allen key 10mm.
10. Use a level to ensure the column assembly is Levelled vertically. Refer
Section "Leveling the ceiling
Mount system" to correct any leveling concerns.
11. After levelling the column. Tighten all 6 bolts to the full torque.
12. Completing of all the procedure from 1 to 11 above is the responsibility
of the customer. Labomed representative will complete rest of the micro-
scope Installation after this.
13. Power up the system with the cable provided. Make sure AC outlet is
hospital grade & grounded properly.
14. Put the drop ceiling tile back into place.
6137000-795
Prima
Wooden membrane
Column
Level
Normal Ceiling Mount
Fig.10
Issue 1.9
Printed on Sept,2021
Prima
Joist
Master Plate
Shim
12

Advertisement

Table of Contents
loading

This manual is also suitable for:

Prima ent6138000-0006137000-000

Table of Contents