23. THERMAL CUT-OFF

The instrument is designed with safety provisions. fans in the electrical box provide free - and forced -
air circulation to cool the electronic components. The instruments also includes a built-in-safety
mechanism called "auto thermal cut-off". This mechanism is activated through Thermistors when the
LED temperature rises above 80 degree C. This time Thermistor sensor activates and provide resistance
to circuit inflow. This causes LED light blinking as an indicator of high LED temperature. Allow the LED
to cool down before switching it on again.

24. TENSION ADJUSTMENT

After
Supplementary
the additional load of suspension arm must be
Compensated by adjusting tension on tension
control screw provided on suspension arm by
moving it clockwise or anticlockwise.
Refer fig.-61
1. Remove plate by unscrewing two holes.
2. Loosen two allen bolts by using allen wrench of
4 mm as shown in fig.62.
3. Use hexagonal wrench of 8mm in bolt (A) shown in
fig.63 rotate it clockwise to increase desired
tension on Gas spring.
4. Re tighten the two screws.
5. Put back the plate.
6137000-795
accessories
are
mounted,
Prima
Prima
Fig. 61
Fig. 62
A
Fig. 63
53
Issue 1.9
Printed on Sept,2021