Installation
(continued)

5.3.2 Positioning on the ceiling

Push the equalising rod with the ceiling plate ④ upward towards the ceiling and affix it
lightly with the four wedges ⑨.
If there is a false ceiling, now draw the cutout ⑤ with Ø 180 mm and saw it out.
If the distance between ceiling and false ceiling is large in the area in which you work,
disassemble the false ceiling elements to provide for freedom of movement during the
installation in the ceiling area.
In an installation without a false ceiling, position the ceiling plate ④ on the ceiling
appropriately such that one of the 3 fastening straps ① faces in the 3 o'clock position. Then
draw the fastening holes.
10
Without false ceiling/with false ceiling
12
Axius
1
2
4
6
3
7
8
9
Part No: 6137120-795
Issue: 1.0
Printed On: June 10, 2023