Table 4
For NON-LIFE-SUPPORT MEDICAL-ELECTRICAL (ME) DEVICES
and ME SYSTEMS
Recommended separation distances between portable and mobile RF
communications equipment and the Veroval
monitor
The Veroval
duo control blood pressure monitor is intended for use in one of the
®
electromagnetic environments specified below in which radiated RF disturbances are
controlled. The customer or user of the Veroval
help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Veroval
control blood pressure monitor, as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation in the
relevant column, whereby P is the maximum nominal power of the transmitter in watts
(W) according to the transmitter manufacturer's specifications.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918.indd 39
®
Separation distance (m) according to transmitter frequency
150 kHz to 80 MHz
80 MHz to 800 MHz
0.12
0.12
0.37
0.37
1.2
1.2
3.7
3.7
12
12
39
English
duo control blood pressure
®
duo control blood pressure monitor can
800 MHz to 2.5 GHz
0.23
0.74
2.3
7.4
23
EN