Do not take your blood pressure after taking a bath or playing sport.
Do not eat, drink or exercise for at least 30 minutes prior to measurement.
Please wait for at least one minute between two measurements.
5.4 Starting the measurement
Start a measurement only after applying the cuff because the cuff may
otherwise be damaged by the excess pressure that is created. Press the
START/STOP button . The appearance of all display segments followed by
the time and date indicates that the monitor is carrying out its self-test
and is ready for use.
Check the display segments for completion.
After about 3 seconds the cuff inflates automatically. If this inflation pressure
is insufficient or if interference is experienced during measurement, the
device will increase pressure to the appropriate higher value in increments
of 30 mmHg. During inflation, the results indicator on the left of the display
also increases at the same time. (If no pulse is detected, the device inflates to
a maximum of about 180 mmHg.)
If a higher inflation pressure is required in principle, you can bypass
the re-pumping by holding down the blue START/STOP button
after the inflation process has started until the desired cuff pressure
is reached. This should be about 30 mmHg above the systolic (upper)
value.
IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918.indd 21
21
English
EN
again
17.10.18 18:03