Hartmann Veroval DUO CONTROL Instructions For Use Manual page 37

Upper arm blood pressure monitor
Hide thumbs Also See for Veroval DUO CONTROL:
Table of Contents

Advertisement

Available languages

Available languages

Electromagnetic in-
terference immunity
tests
Magnetic field at
supply frequency
(50 Hz) as per
IEC 61000-4-8
Table 3
For NON-LIFE-SUPPORT MEDICAL-ELECTRICAL (ME) DEVICES and ME
SYSTEMS
Manufacturer declaration and guidance – Electromagnetic immunity
The Veroval
duo control blood pressure monitor is intended for use in one of the
®
electromagnetic environments specified below. The customer or user of the Veroval
control blood pressure monitor should assure that it is used in such an environment.
Electromagnetic
interference immunity
tests
conducted RF
disturbances as per
IEC 61000-4-6
radiated RF interference
in accordance with
IEC 61000-4-3
IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918.indd 37
IEC 60601 test
Compliance level
level
3 A/m
3 A/m
IEC 60601 test level
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to
2.5 GHz
37
Electromagnetic environment
– guidance
Power frequency magnetic
fields should be at levels
characteristic of those found
in a typical commercial or
hospital environment.
Compliance
Electromagnetic environment
level
– guidance
Portable and mobile RF
communications equipment
should be used no closer
to any part of the Veroval
duo control blood pressure
monitor, including its cables,
than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter. Recommended
separation distance:
3 V
3 V/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
English
EN
duo
®
®
17.10.18 18:04