Installation
5.3.3
Connecting 2-wire temperature sensors
Disconnect the CALEC® ST III from the power
u
supply.
Take off the front cover [}p. 20].
u
Connect the temperature sensor:
u
T
terminals 1/2
l
hot
T
terminals 3/4
l
cold
Connect the four wire bridges:
u
Between terminals 5/1
l
Between terminals 2/6
l
Between terminals 7/3
l
Between terminals 4/8
l
Close the front cover [}p. 20].
u
5.3.4
Connecting 4-wire temperature sensors
Disconnect the CALEC® ST III from the power
u
supply.
Take off the front cover [}p. 20].
u
Connect the temperature sensor:
u
T
terminals 1/5 and 2/6
l
hot
T
terminals 3/7 and 4/8
l
cold
Close the front cover [}p. 20].
u
26 / 56
Incomplete wiring of the temperature sensors
Incomplete wiring of the 2-wire temperature sensors may lead to false measurement results.
Always connect the 2-wire temperature sensors together with the four supplied wire
u
bridges.
NOTICE
3-140-UM-EN-02