3.0 Setting up and starting up
3.2 Assembly
Please make sure that the static conditions stated by ATMOS MedizinTechnik are met (for details, see the separately enclosed
!
document "Static requirements for installing the ATMOS i View"). The fulfilment of these requirements must be confirmed by an
authorized expert.
Mains voltage and fuse: Mains voltage: 100–240 V, 50/60 Hz; Fuse: 2 x T 3.15 A
Please note that only PCs and monitors with IEC 60601-1/EN 60601-1 approval may be connected to the video outlets of the
ATMOS i View COLPO supply module!
3.2.1
Connection to the power supply
Potential equalization:
The ATMOS i View COLPO's supply module has a rear connection for potential equalization which can be connected to the poten-
tial equalization rail in the room if need be. Hereby, user/patient safety can be increased especially in the case of a defective earth
conductor. For connecting the device's potential equalization plug with the potential equalization rail of the room, use the potential
equalization cord with REF 530.0030.0.
3.2.2
Colposcope overview
Wide-field eyepiece
lenses with binocu-
lar tube
T-hand grip (optional)
10
Brightness control
Lateral double hand
grip (optional)
Control panel
(optional)
Rotating disk for
5-fold magnification
changer
Lens
Pivoting H.A.S.I. color filter (optional)
Measuring scale (optional)
HD adapter for SONY
digital camera (optional)