Disinfection:
Purpose: To destroy or inactivate micro-organisms
• Use gloves for this operation
• Apply a disinfectant with a carrier moistened with a disinfectant solution
• Rinse thoroughly with water and a sponge
Drying:
Purpose: Protect equipment from contamination once disinfected
• Wipe the equipment with a dry, lint-free clean
Storage:
Objective: To maintain the cleanliness of the clean and disinfected devices
• Store equipment and medical devices, once retired, in clean places, raised above ground and protected from dust,
moisture, heat, sunlight and risk of deterioration of packagings (tears, staples, ...)
Product should be avoided for cleaning and disinfecting:
• peracetic acid at concentrations> 10%
• organic solvents like halogenated hydrocarbons / aromatic or acetone
• products with a pH> 8 and <5
• products containing corrosive or caustic substances
Recommended products for cleaning and disinfecting:
• Argos spray 750 ml Argos 1861
• Neodisher® Dekonta, Neoform MED AF,Neoform MED FF,Neoform MED Spray
• SEPTOTAL S-911
• Surfanios
• Axiol
In the case of use of a specialized product, we strongly advise to adhere to the protocol recommended by
the manufacturer.