<Attachment>
CAUTION
Slowly insert the air/water valve straight into the air/water valve cylinder of the endoscope.
If the air/water valve is attached forcibly, it may be damaged.
Note
The air/water valve and the air/water valve cylinder have a blue marking. Check the
marking on the air/water valve and that on the air/water valve cylinder to prevent an error
in attachment.
(1) Attach the air/water valve to the endoscope's air/water valve cylinder, and push in the valve
firmly.
(2) Visually and manually check that the valve fits properly without any bulging of the skirt.
897N201625
Blue marking
Blue marking
4.4 Inspecting and Attaching Accessories
Forceps Valve Suction Valve
Air/Water Valve
Bulge
J Tube