FujiFilm 700 Series Operation Manual page 138

Lower gastrointestinal endoscopes
Table of Contents

Advertisement

Appendix
Immunity test
Immunity to
380 - 390 MHz, 27 V/m
proximity fields
430 - 470 MHz, 28 V/m
from RF wireless
704 - 787 MHz, 9 V/m
communications
800 - 960 MHz, 28 V/m
equipment
1422 - 1512 MHz, 10 V/m
IEC/EN 61000-4-3
1700 - 1990 MHz, 28 V/m
2400 - 2570 MHz, 28 V/m
3480 - 3600 MHz, 10 V/m
5100 - 5800 MHz, 9 V/m
Note
 At 80 MHz and 800 MHz, the higher frequency range applies.
 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which this
product is used exceeds the applicable RF compliance level above, this product should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating this product.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
c. Frequency bands of 6.765 MHz to 6.795 MHz, 13.553 MHz to 13.567 MHz, 26.957 MHz to 27.283 MHz,
or 40.66 MHz to 40.70 MHz
Appendix-12
IEC 60601
Compliance level
Test level
380 - 390 MHz, 27 V/m
430 - 470 MHz, 28 V/m
704 - 787 MHz, 9 V/m
800 - 960 MHz, 28 V/m
1422 - 1512 MHz, 10 V/m
1700 - 1990 MHz, 28 V/m
2400 - 2570 MHz, 28 V/m
3480 - 3600 MHz, 10 V/m
5100 - 5800 MHz, 9 V/m
Guidance
Degradation of the performance
of this product could result if
portable RF communications
equipment is used closer than 30
cm to any part of this product.
897N201625

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ec-760z-v/mEc-760z-v/lEc760zpvl

Table of Contents