<Electromagnetic Immunity Compliance Information and Guidance>
Table 2
Guidance and Manufacturer Declaration - Electromagnetic Immunity -
This device is intended for use in the following prescribed electromagnetic environments. Customers and
users of this product are advised to check that it is being used in such environments.
Immunity test
Electrostatic discharge
(ESD)
IEC/EN 61000-4-2
Electrical fast transient/
burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC/EN 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC/EN 61000-4-8
Note
 U
is the a.c. mains voltage prior to application of the test level.
T
Appendix-10
IEC 60601
Test level
Distal end:
Distal end:
±6 kV contact
±6 kV contact
±2 kV, ±4 kV, ±8 kV air
±2 kV, ±4 kV, ±8 kV air
Other parts:
Other parts:
±8 kV contact
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±15 kV air
±2 kV for power supply
±2 kV for power supply
lines
lines
±1 kV for input/output
±1 kV for input/output
lines
lines
±0.5 kV, ±1.0 kV line to
±0.5 kV, ±1.0 kV line to
line
line
±0.5 kV, ±1.0 kV,
±0.5 kV, ±1.0 kV,
±2.0 kV line to earth
±2.0 kV line to earth
0%U
0%U
T
for 0.5 cycles and
for 0.5 cycles and
1 cycles
1 cycles
70%U
70%U
T
for 0.5 second
for 0.5 second
0%U
0%U
T
for 5 seconds
for 5 seconds
30 A/m
30 A/m
Compliance level
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
T
that of a typical commercial or
hospital environment. If the user
of this product requires continued
operation during power mains
T
interruptions, it is recommended
that this product is powered from
an uninterruptible power supply
T
or battery.
The power frequency magnetic
field should have the same level
of characteristics as a common
location in standard business
and hospital environments.
Guidance
897N201625