<Electromagnetic Immunity Compliance Information and Guidance>
Table 6
Guidance and Manufacturer Declaration - Electromagnetic Immunity -
This device is intended for use in the following prescribed electromagnetic environments. Customers and
users of this product are advised to check that it is being used in such environments.
Immunity test
Electrostatic discharge
(ESD)
IEC/EN 61000-4-2
Electrical fast transient/
burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC/EN 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC/EN 61000-4-8
Note
 U
is the a.c. mains voltage prior to application of the test level.
T
897N201625
IEC 60601
Test level
±2 kV contact
±2 kV contact
±4 kV contact
±4 kV contact
±6 kV contact
±6 kV contact
±2 kV air
±2 kV air
±4 kV air
±4 kV air
±8 kV air
±8 kV air
±2 kV for power supply
±2 kV for power supply
lines
lines
±1 kV for input/output
±1 kV for input/output
lines
lines
±0.5 kV, ±1.0 kV line to
±0.5 kV, ±1.0 kV line to
line
line
±0.5 kV, ±1.0 kV,
±0.5 kV, ±1.0 kV,
±2.0 kV line to earth
±2.0 kV line to earth
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5 cycle
for 0.5 cycle
40%U
40%U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5 s
for 5 s
3 A/m
3 A/m
Compliance level
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Main power quality should be
that of a typical commercial or
hospital environment.
Main power quality should be
that of a typical commercial or
hospital environment.
T
)
T
Main power quality should be
that of a typical commercial or
T
)
hospital environment. If the user
T
of this product requires continued
operation during power mains
interruptions, it is recommended
T
that this product is powered from
)
T
an uninterruptible power supply
or battery.
T
)
T
The power frequency magnetic
field should have the same level
of characteristics as a common
location in standard business
and hospital environments.
Guidance
Appendix-15