Chapter 1 Precautions
CAUTION
 Turn off the light of the light source except during an inspection, procedure, etc., when
necessary. If the light of the light source is left on, the distal end of the endoscope and its
surroundings may become hot, causing burn injury to the patient or end-user.
 When turning off the processor, also turn off the light source. If the light source remains on
after turning off the processor, the ALC (automatic light control) does not function and the
maximum amount of light is emitted. As a result, the distal end of the endoscope and its
surroundings may become hot, causing burn injury to the patient or end-user.
 Use this product only in combination with related equipment described in this manual.
Otherwise, it creates a risk of equipment malfunction.
 Use the endoscopic CO
connected, the air/water supply function lessens and may result in improper cleaning of the
lens.
 Set the suction pressure between 40 and 53 kPa. If the suction pressure is too high, the
endoscope may adhere to mucous membrane, resulting in damage to the mucous
membrane.
 Do not subject the scope connector to strong impact. Install the light source away from
obstacles to prevent the scope connector connected to the light source from accidental
impact damage. During the operation of an electric bed, etc., ensure that the scope
connector connected to the light source does not hit the bed. Otherwise, the scope
connector of the endoscope and the light source may malfunction.
 With regard to the amount of sterile water in the water tank, follow the instructions provided
in the operation manual of the water tank. If the amount of sterile water in the water tank
exceeds the limit, the air/water supply function may be disabled or it may cause equipment
failure due to contact with leaked sterile water.
 Attach the water tank to the specified position of the cart or light source. Otherwise, fluid
may leak from the connector of the water tank and it may come into contact with related
equipment, causing equipment failure.
 Firmly connect the suction tube from the suction unit to the suction connector on the scope
connector. If the suction tube is not attached properly, body fluid may drip from the tube and
come into contact with related equipment, causing equipment failure.
Note
 For details on how to use related equipment, refer to the operation manual of related
equipment.
 Before using this product, thoroughly read the operation manual of related equipment
used in combination with this product.