FujiFilm 700 Series Operation Manual page 137

Lower gastrointestinal endoscopes
Table of Contents

Advertisement

<Portable and Mobile RF Communications Equipment Compliance Information and
Guidance>
Table 3
Guidance and Manufacturer Declaration - Electromagnetic Immunity -
This device is intended for use in the following prescribed electromagnetic environments. Customers and
users of this product are advised to check that it is being used in such environments.
Immunity test
Conducted RF
3 Vrms
IEC/EN 61000-4-6
150 kHz to 80 MHz
6 Vrms
ISM Frequency Band
Radiated RF
3 V/m
IEC/EN 61000-4-3
80 MHz to 2.7 GHz
897N201625
IEC 60601
Compliance level
Test level
3 Vrms
150 kHz to 80 MHz
6 Vrms
c
ISM Frequency Band
3 V/m
Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of this product, including
cables, than the recommended
c
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
d = 1.2
√ P
d = 1.2
80 MHz to 800 MHz
√ P
d = 2.3
800 MHz to 2.7 GHz
√ P
Where "P" is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and "d"
is the recommended separation
distance in metres (m). The
electric field intensity from a fixed
RF transmitter determined by an
electromagnetic field study
should be lower than the
compliance level in each
b
frequency range
.
Interference may occur near
devices that display the following
symbol.
a
Appendix-11