FujiFilm 700 Series Operation Manual page 23

Lower gastrointestinal endoscopes
Table of Contents

Advertisement

WARNING
 Never use endotherapy devices, operate, insert or withdraw the endoscope, or operate the
related equipment without viewing or while freezing the endoscopic image on the monitor.
Otherwise, injury to tissues in the body cavity, bleeding and/or perforation may result.
 Do not use endotherapy devices, operate, insert or withdraw the endoscope, perform
bending, air supply, suction, water jet operation or flexibility adjustment, or operate the
related equipment whenever the endoscopic image is compromised, unclear, blurry, etc.
due to any reason or condition including loss of image, power interruption, water droplets or
dirt/debris adhering to the objective lens, etc. Doing so may cause injury to tissues in the
body cavity, bleeding and/or perforation.
 If a patient sneezes or moves abruptly during the procedure, malfunction of the endoscope
and patient bleeding or trauma may occur. Depending on the degree of malfunction, safe
endoscope withdrawal may be difficult or impossible, causing severe harm to patient and/or
end-users.
 Do not perform retroflexed observation forcibly. Performing retroflexed observation in a
narrow lumen may make it impossible to straighten the angle of the bending section and/or
withdraw the endoscope from the patient.
 When this product is used for a patient with an active implantable medical device such as a
pacemaker, consult a cardiovascular specialist and the manufacturer of the active
implantable medical device to ensure patient safety. The radio waves radiated from this
product may cause medical devices such as a pacemaker to malfunction or break down,
seriously affecting patient safety.
 Do not look directly into the light coming from the light guide at the distal end of the
endoscope. Viewing the light from the light guide directly may damage your eyes.
 Do not apply excessive force of the endoscope or endotherapy device against mucosal
surfaces. Doing so may cause injury to tissues in the body cavity, bleeding and/or
perforation.
 Do not bend or insert the endoscope while an endotherapy device protrudes from the distal
end. Excessive force of the endotherapy device may be unintentionally applied against
mucosal surfaces, causing injury to tissues in the body cavity, bleeding and/or perforation.
 Use the air/water channel cleaning adapter only for pre-cleaning of the air/water channel. If
it is used during a procedure, continuous air supply may occur and cause patient injury.
 Do not forcibly advance or withdraw the endoscope into/from the patient, angulate the
bending section forcibly or operate it quickly. It may cause patient injury, bleeding or
perforation.
 Insert the endoscope while observing the endoscopic image to secure patient safety. Not
doing so may cause patient pain, injury, bleeding and/or perforation.
 When using special light observation mode, use it after sufficiently understanding the
difference between the color tone and brightness of normal light observation mode and
those of special light observation mode. Use images displayed in special light observation
mode as reference information only. Always use the standard endoscopic image in normal
light observation mode when performing a diagnosis. Otherwise, it may cause
misdiagnosis.