9.5
Return
The following procedures must be carried out before a transmitter is sent to Endress+Hauser e.g.
for repair or calibration:
• Remove all residue which may be present. Pay special attention to the gasket grooves and crevices
where fluid may be present. This is especially important if the fluid is dangerous to health, e.g.
corrosive, poisonous, carcinogenic, radioactive, etc.
• Always enclose a duly completed "Declaration of contamination" form (a copy of the "Declaration
of contamination" is included at the end of this operating manual). Only then can Endress
+Hauser transport, examine and repair a returned device.
• Enclose special handling instructions if necessary, for example a safety data sheet as per EN 91/
155/EEC.
Additionally specify:
• An exact description of the application
• The chemical and physical characteristics of the product
• A short description of the error that occurred (specify error code if possible)
• If necessary, give the error code
9.6
Disposal
In case of disposal please seperate the different components according to their material consistence.
9.7
Software history
Software modifications
Original software.
Operated via:
– ToF Tool
– Commuwin II (as of Version 2.08-1 Update C)
– HART-Communicator DXR375 withRev. 1, DD 1.
• "mapping" function improved
Function "echo lost" improved
• function "detection window"
"Echo" and "linearisation" improved
Function "LAS" improved
Documentation
Description of Device
Functions
BA281F/00/en/04.04
52021043
BA281F/00/en/06.04
BA245F/00/en/06.04
52021043
52011936
BA281F/00/en/04.05
52021043
BA281F/00/en/01.06
BA245F/00/en/01.06
52021043
52011936
-----
BA281F/00/en/05.06
52021043
BA245F/00/en/06.06
52011936
BA281F/00/en/11.06
52021043
BA281F/00/en/12.06
52021043
-----
BA281F/00/en/03.09
71074933
Trouble-shooting
-----
----
-----
-----
87
Need help?
Do you have a question about the Fieldbus Levelflex M FMP45 and is the answer not in the manual?