Commissioning
⇒
ENDRESS + HAUSER
–
+
E
⇒
ENDRESS + HAUSER
–
+
E
!
58
Function "dist./meas.value" (008)
The distance measured from the reference point to the product surface and the measured value
calculated with the aid of the empty calibration are displayed again. Check whether the values
correspond to the actual meas. value or the actual distance. The following cases can occur:
• Distance correct – meas. value correct -> basic setup completed
• Distance incorrect – meas. value incorrect -> another interference echo suppression must be
carried out "check distance" (051).
• Distance correct – meas. value incorrect -> check "empty calibr" (005)
⇓
Note!
After the basic setup, it is recommended to evaluate the measurement with the aid of the envelope
curve ("envelope curve" (0E) function group) (→ ä 61).
Return to group selection
After the interference echo suppression the basic setting is finished and the instrument jumps
automatically back into the group selection.
Levelflex M FMP45 with FOUNDATION Fieldbus