Mixtures; Valves; Pumps - Tecfluid FLOMID Series Instruction Manual

Sensor and converter
Hide thumbs Also See for FLOMID Series:
Table of Contents

Advertisement

4.2.1

Mixtures

If liquids of different conductivities are mixed it is necessary to install the sensor a minimum
of 30 DN from the point of mixture in order to obtain a uniform conductivity of the liquid and
stabilize the readings.
If this distance is shorter, readings may be unstable.
4.3

Valves

Control valves or shut-off valves should always be installed downstream from the sensor to
assure that the pipe is always full of liquid.
4.4

Pumps

Pumps should be mounted upstream from the sensor to avoid the suction of the pump
(vacuum) that could damage the sensor liner.