c)
Installation in which for a time a first product flows through the pipe, and after that a
second product flows in the opposite direction. The programming will be:
Positive flow rate action: Increase
Negative flow rate action: Increase
The totalizer will indicate the volume corresponding to the sum of both products.
Note:
To reset the totalizer, go to chapter 5 (page 27).
7.8

Modbus

This option is only implemented in some devices.
It allows to configure the needed parameters to establish a Modbus RTU communication
with the instrument.
Detailed information on this point could be found in the R-IT-MX4COM instructions manual,
that can be downloaded from the Tecfluid S.A. website.
8

SERIAL NUMBER

In this section the converter serial number is shown.
9

SOFTWARE VERSION

From the main menu, if "Software version" is selected, the screen will show this data and the
corresponding date.
10

WORKING SCREEN

When exiting the menu, the display shows the default screen. To scroll between the three
operating screens, press the key (Down / Left) or (Up).