Table of Contents

Advertisement

10 Technical data

10.4 Fluidics

Item
Fluid system
Flow cuvette
Sample delivery
Sample volume
Flow rates
Fluids volume
Bio safety system
42
Specification
Completely closed system for Sheath Fluid and sample
volumes
No fluid droplets or aerosols generated nor released from
the device
The pressure of the closed system is below 500 mbar
Synthetic quartz flow cuvette with centric flow channel
(capillary diameter 350 μm × 200 μm) for laminar sample
transport with Sheath Fluid
Computer controlled precision syringe pump for
contamination-free sample transport.
Built-in vacuum pump for Waste bottle; vacuum pressure
is adjustable (computer controlled)
Continuous up to 1200 μl
Minimal content of sample volume 850 μl, an additional
dead volume of 50 μl is recommended for best results
200 μl for electrode-based precision absolute counting
(other counting volumes upon request)
Minimal sample volume 10 μl
10 μl to 1000 μl for syringe-based precision absolute
counting
Sample volume speed adjustable continuously between
0.1 μl/s and 19,9 μl/s)
2 glass bottles (each 1 litre) for Sheath Fluid and waste
Avoids sample droplets and sample cross contamination
(computer controlled)
CyFlow™ Cube 6 V2m | Operating Manual | February 2020

Advertisement

Table of Contents
loading

Table of Contents