11.5.5 Acquisition

Multiple trigger parameters are logical "OR" connections.
Property
Light Source
Trigger Parameter
Gain Settings
Tresholds
Fluidics
Flow Rate
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
Figure 31: Acquisition register
Description
Flags "Laser" [...] X selects the active light sources in the device
Flags "FSC", "SSC", [...], "FL4" select trigger parameters
The bars set the PMT signal amplification for the individual
optical parameter in voltage (0 – 999 V)
The bars set the cut off level for trigger signals
(left-click shows the cut off level in V)
Sets the speed of sample flow in µl/s