Electrical Hazards; Alterations To The Device - SYSMEX CyFlow Cube 6 V2m Operating Manual

Table of Contents

Advertisement

Only authorised service personnel is allowed to open the housing of the
device.
Laser light of laser class IIIb can be emitted if the housing of the device is
damaged or the laser protection is removed. Serious injuries of the eyes
and the skin are the results of emission by laser light of class IIIb lasers.

2.10 Electrical hazards

The main electrical source used must be grounded or earthed due to local
specifications. The mains cable carries out the protective earthing. Please
place the device in such a way that the electrical power cord can easily be
unplugged. Removable power cords should not be replaced by inadequate
wiring.
Beware of possible voltage when changing the fuse (T 3.15 A 250 V).

2.11 Alterations to the device

Unauthorised alterations to the device can result in risks and hazards. Therefore, unless
expressly permitted by the manufacturer, no alterations may be made to the device.
If the device is altered, appropriate tests and trials must be performed to ensure the
continued safe use of the device.
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
2 Safety