11 Software
Measure Mode
VC Electrodes
VC of Volume
Analyze all
Cells in Region
Parameters of Workstep
NOTICE
Sample Flow
It is recommended to choose the same sample flow for pre-run and
measure to avoid fluid turbulence.
Speed and Volume values can be stored by saving a configuration file.
Property
Options
Command +
Flow
Parameter
Volume
62
Table 17: Manual Sample Port Measure Modes
Description
True Volumetric Absolute Counting. The start and stop
electrode are used to measure a defined volume of 200 µl.
An adjustable predefined volume is analysed.
Continuous acquisition of the complete sample.
Continous acquisition until a flexible predefined number of
particles within in a specified region is reached.
Figure 34: Workstep register
Description
Adjusts the sample flow [µl/s] for pre-run and
measure phases
Adjusts the measured volume [µl] for pre-run and
measure phases
CyFlow™ Cube 6 V2m | Operating Manual | February 2020