11. Measuring menus
If the temperature of the medium deviates from 25°C the following measuring
errors must be considered :
e.g. Errors per 10 °C
Dynamic pressure
pH probe
approx. 3.3%
For temperature compensation of these sensors there are 2 possibilities :
Enter compensation temperature
in function
Both in this function and in the measured value the symbol
Continuous temperature compensation with external temperature sensors
can be activated either via the reference channel (see 13.12.6) of the sensor to
be compensated or by configuring any temperature sensor as reference sensor
with a ´*T´ in the designation (see 13.2).
While the temperature is being measured, a point flashes
behind the symbol ´CT´
Flow values (velocity or volume flow) acquired with temperature com-
pensation can by means of ´#N´ in the designation (see 13.2) be con-
verted to standard conditions of 20°C (see Manual 6.7.5).
11.2.5
Measured variables dependent on the ambient atmospheric pressure (see Sec-
tion 13.9 Measuring range list ´with PC´) may, in the event of large deviations
from normal pressure (1013 mbar), involve certain measuring errors.
e.g. Error per 100 mbar
Rel.humidity psychrometer
Mixture ratio, capacitive
Dynamic pressure
O
saturation
2
It is advisable therefore, especially when taking measurements at appreciable
heights above sea level to take due account of the atmospheric pressure (ap-
prox. -11 mbar / 100 meters above mean sea level, MSL). On all sensors requir-
ing atmospheric pressure compensation the
atmospheric pressure
The appropriate atmospheric pressure can be entered either in
or in device programming (see 14.6) or it can be measured using an atmos-
pheric pressure sensor (reference sensor with designation ´*P´, see 13.2, Man-
ual 6.7.2). While the atmospheric pressure is being used for compensation both
in the function
appears; if it is being measured a point flashes behind
CP
Please note that as soon as a reference sensor is disconnected nor-
mal pressure, 1013 mbar, is used.
Flow values (also volume flow with rotating vanes) acquired with at-
30
Compensation range
approx. 1.6%
0 to 100 °C

Atmospheric pressure compensation

Compensation range
approx. 2%
approx. 10 %
approx. 5%
approx. 10%
:
atmospheric pressure
Temp. comp. CT
500 to 1500 mbar
Vapor pressure VP up to 8 bar
800 to 1250 mbar (error < 2%)
500 to 1500 mbar
sensor display includes
atm pressure: CP
and at the measured value the symbol
Sensor
-50 to 700 °C NiCr-Ni
Ntc or Pt100
appears.
´CT´
temp. comp. CT.
1013 mbar.
sensor display
CP.
31.0°C
23.5°C
the function