Two-Point Adjustment With Setpoint Entry - Ahlborn ALMEMO 2590-2A Operating Instructions Manual

Universal measuring instruments and data loggers
Table of Contents

Advertisement

To clear all measured values, if necessary, press key
The average value of the array is displayed continuously:
To cancel the measuring operation press key :
To proceed to the 3rd menu press key :
Display of volume flow in m
It is possible to display the standardized volume
referred to 20°C and 1013mbar if the temperature
and atmospheric pressure are measured at the
measuring point in question; (see temperature
compensation 11.2.4 and atmospheric pressure
compensation 11.2.5).
Select the function standardized and activate :
Save data and output
If a SD-card memory connector connected, all data can be saved to a special vol-
ume flow file by pressing key :
The tabular format of the data is shown on the previous page. The data can
command 'P49' (s. manual sec. 6) can also be read through the interface, and
then imported for processing in an Excel spreadsheet.

12.3 Two-point adjustment with setpoint entry

For universal error correction at any 2 points the
function menu
two-point adjustment
If the actual values at 2 points are known , these
can be written with the appropriate setpoints. If
not, 2 setpoint states must be created and ad-
justed online. Usually, for the 1st measuring point,
a zero-point adjustment is performed; however, any
other
setpoint is equally possible. For the 2nd measuring point, a gain adjustment i per-
formed and all correction values are recalculated; (see 13.7).
Two-point adjustment : (actual values are cleared)
1. Measuring point
Place the sensor in 1st status
(e.g. icy water, unpressurized, etc.),
Select setpoint 1 and enter :
Adjust measured value to setpoint 1 by pressing :
Measured value should now display setpoint 1 : 00:
2. Measuring point
Place the sensor in 2nd status
(boiling water, known weight, etc.)
For the 2nd measuring point enter setpoint 2 :
Adjust gain in function setpoint 2 by pressing :
Measured value should now display setpoint 2 : 00:
Correction value calculation
/h = Average value of flow [m/s] x cross-section [cm
3
is provided.
12. Function menus
< CLRA >
average val : 15.11 m¡
< ESC >
< VOL ►>
standardized :
<STORE>
00:
0.4 °C
actual value 1: -----
setpoint 1:
0.0
<ADJ>
0.0 °C
00:
99.45 °C
<ADJ>
100.0 °C
]
2
2: -----
2:
100.0
43