11. Measuring menus
11.2.1
One very useful function is to zero the measured
value at certain locations or at certain times as a
reference value in order then to observe only the
subsequent deviations. Having selected the
measured value function the softkey
appear. Press this key to save the measured
value displayed as base value (see 13.6) and re-
set to zero.
Select measured value function (see 9.4) :
To select the zero-set measured value function press
The measured value will then show :
The base value is then assigned the measured value :
To cancel zero-set, then after pressing :
If the function is locked (see 13.4) the base value is not saved on the
connector but only temporarily to RAM where it is retained until the
device is next switched off. This status is indicated in the display by
the symbol
If you prefer to disable the zero-set function completely, the channel
in question must be locked at level 6.
11.2.2
Dynamic pressure probes FDA602Sx must un-
dergo zero-point adjustment before each measur-
ing operation by withdrawing the hoses. The zero-
point error is always written to the calibration off-
set temporarily, i.e. until the device is next
switched off, irrespective of locking level - thus en-
suring that linearization is not falsified.
To select the measured value function press key(s) :
To perform zero-point adjustment press key(s) :
11.2.3
probes
The following chemical sensors need to be ad-
justed at least once or at regular intervals to com-
pensate for various instabilities. In the measured
value function pressing key
automatically perform a two-point adjustment of
the zero-point and the gain. As and when adjust-
ment is performed the appropriate calibration set-
points appear; these can also be modified as re-
quired:
28

Set measured value to zero

; in other cases the symbol º appears.
REL

Sensor adjustment for dynamic pressure probes

Sensor adjustment for chemical sensors and
<ADJ>
will
<ZERO>
00:
00: 00.0 °C and the symbol REL
<ZERO>
can be used to
00:
23.4
NiCr Temperatur
ZERO ESC F
23.4
°C
<ZERO>.
base value: 23.4 °C
hold the key down.
PROG
... (see 9.4).
<ADJ>
°C
REL
.
.