Using List Frequency Mode - Keysight Technologies N9040B UXA Manual

Signal analyzers
Hide thumbs Also See for N9040B UXA:
Table of Contents

Advertisement

Measurement Related Tasks and Concepts
Setting the Signal Analyzer to Make a Measurement

Using List Frequency Mode

List frequency mode allows you to enter the frequency points where
measurements are made. This allows you to specify measurement points, for
example, in areas of interest that would otherwise have less coverage in the
sweep mode.
Frequency lists are limited to 501 entries.
To set the analyzer to use the data in the frequency list table, press FREQ,
Freq Mode, List.
You can create a frequency list in the following ways:
— Manually, by specifying each individual point.
— From the sweep points, by specifying the measurement frequency range
— Loading a list from the internal memory where the data has been previously
To creating a frequency list manually,
Step
FREQ
Freq Mode
1. Press
,
Ed it Freq List
2. Press
appears.
N9069C Noise Figure Measurement Guide
and setting the analyzer to generate equally spaced points within that
range, press Ed it Freq List, enter the start and stop frequencies, press Fill
Using Start and Stop Freq key. This list of frequencies can be edited later if
required.
stored.
List
,
.
Ed it
,
. A Frequency List table
Notes
71

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents