Measurement Related Tasks and Concepts
Setting the Signal Analyzer to Make a Measurement
— When in noise figure mode, SNS ENR data by default will load
The 346 Series noise sources from Keysight Technologies have the ENR
values printed on a label that is affixed to the body of the device. These
ENR values are also provided in the form of a calibration report, and on a
diskette which is supplied with all Keysight 346x Series noise sources. The
values printed on the noise source itself are only shown to two decimal
places. The values stored on a diskette are correct to three decimal places.
To enter the ENR data manually, do the following:
Table 3-2
Step
1. Power on the analyzer and wait the recommended
time for warmup.
MODE/MEAS
2. Press
MEAS SETUP
3. Press
Figure 3-4
66
automatically when the SNS is connected to the analyzer.
Noise Figure, OK
,
.
ENR
,
.
ENR Dialog
Notes
N9069C Noise Figure Measurement Guide