Configuring Fixed Loss Compensation - Keysight Technologies N9040B UXA Manual

Signal analyzers
Hide thumbs Also See for N9040B UXA:
Table of Contents

Advertisement

Measurement Related Tasks and Concepts
Setting the Signal Analyzer to Make a Measurement

Configuring Fixed Loss Compensation

To configure fixed loss compensation follow the example below:
Step
MEAS SETUP
1. Press
Loss, Before DUT
2. Press
Fixed
Comp Mode,
3. Press Fixed Loss and enter the loss
compensation value and press the
termination key.
Temperature
4. Press
keys or the knob to enter the temperature of
the device where the loss is occurring. This
will normally be room temperature, which is
290 K.
N9069C Noise Figure Measurement Guide
Loss Comp
,
tab.
After DUT
or
, Loss
.
dB
, and use the numeric
Notes
The lower limit is –100.000 dB, the upper limit is
100.000 dB, and the default is 0.000 dB.
It is important that you enter the correct temperature. This
will give you better accuracy for you measurements.
79