Making Noise Figure Measurements
Making Frequency Converter Measurements
Step
6. Configure Ext LO Setup
7. Configure DUT LO Setup
8. Configure the DUT
Setup
24
Action
Ext LO Setup
— Press
.
LAN
— In the
box of the
IP
dialog, enter the
Address
and press
Done
.
Add Specified IP
— Press
Address
.
Select
— Press
Highlighted Source as
DUT LO
.
— On the Select dialog,
Ext LO Setup
press
.
DUT LO Setup
— Press
LO Settings
— In the
area
of the dialog, enter the
following values:
LO Power; 7.0 dBm
Min Freq; 250 kHz
Max Freq; 6 GHz
.
DUT Setup &
— Press
Calibration
.
— Set the parameters as
Table 2-2
listed in
.
Notes
For more information about setting an external
LO, see
"Setting up External LO Control" on
page
77.
.
This sets the minimum and maximum values of the
signal generator.
After setting all parameters, the DUT Setup screen
will look like the following figure.
"DUT Setup Form" on page 61
Refer to
see detailed information about each parameters in
the DUT setup form.
N9069C Noise Figure Measurement Guide