Making Noise Figure Measurements
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Figure 2-17
Supported DUT Characteristics and Recommended Test System
Configuration for Frequency Ranges above 3.6 GHz
For frequency ranges above 3.6 GHz, if no preamplifier is used in the signal
analyzer, the major noise power concern will be the noise power before the mixer.
In this case, the input noise outside the preselector bandwidth (less than 80 MHz)
will be filtered. This may affect the supported DUT characteristics.
Figure 2-16
26.5 GHz. For the frequency ranges above, the U7227F USB Preamplifier used
together with the X-Series can support up to 50 GHz. The calculation method
for frequency ranges above 26.5 GHz is similar to the "3.6 GHz to 26.5 GHz"
frequency range. The supported DUT gain will decrease because of the
increase of the noise figure and gain for both the USB preamplifier and the
internal preamplifier. Refer to
N9069C Noise Figure Measurement Guide
and
Figure 2-17
provide a rough guide for frequency ranges up to
Table 2-3