4.Using Test Mode
4.4 Test Menu 4: Analog Output Operation Test
The following operations are available in the test mode.
When the ME-4210-SS96 optional plug-in module is not installed, this test menu will be skipped.
Test menu
DISPLAY
SET
Analog output CH1
operation test
DISPLAY
SET
Analog output CH2 to
CH4
operation test
SET
Test menu
Set the test menu number to "4".
Execute the operation test for analog output CH1. (Input is not necessary.)
Output
0%
25%
50%
75%
100%
100%
Note: The analog output item set up with Setting the Analog Output(Setting Menu 6) is displayed.
Execute the operation test for analog output CH2 to CH4.
The test method is the same as ①Analog output CH1 operation test .
<To continue other test mode>
Select other test menu number and then press
<To close test mode>
Select "End" as the test menu number and then press
You will return to the operation mode.
Output Specification
4-20mA
0%
4mA
25%
8mA
50%
12mA
75%
16mA
20mA
SET
- 46 -
Upper :Analog output 1
Middle :% value
Lower :Setting analog output item
and the measurement value