Beckman Coulter Spinchron DLX Instruction Manual page 3

Benchtop centrifuge
Table of Contents

Advertisement

!
SAFETY NOTICE
This safety notice summarizes information basic to the safe operation of the equipment described
in this manual. The international symbol displayed above is a reminder that all safety instructions
should be read and understood before installation, operation, maintenance, or repair of this centri-
fuge. When you see the symbol on other pages, pay special attention to the safety information
presented. Observance of safety precautions will also help to avoid actions that could damage or
adversely affect the performance of the centrifuge.
Safety During Installation and/or Maintenance
This centrifuge weighs 65.4 kg (144.0 lb). DO NOT attempt to lift or move the centrifuge without
assistance from another person.
Be sure to use the anchoring system to secure the centrifuge in place. The anchoring system is
designed to reduce the possibility of injury or damage which could result from centrifuge move-
ment in the event of a major rotor mishap.
Any servicing of this centrifuge that requires removal of any covers can expose parts which involve
the risk of electric shock or personal injury. Make sure that the power switch is off and the centri-
fuge is disconnected from the main power source, and refer such servicing to qualified personnel.
Do not replace any centrifuge components with parts not specified for use on this instrument.
Electrical Safety
To reduce the risk of electrical shock, this centrifuge uses a three-wire electrical cord and plug to
connect this equipment to earth-ground. To preserve this safety feature:
• Make sure that the matching wall outlet receptacle is properly wired and earth-grounded. Check
that the line voltage agrees with the voltage listed on the name-rating plate affixed to the
centrifuge.
• Never use a three-to-two wire plug adapter.
• Never use a two-wire extension cord or a two-wire non-grounding type of multiple-outlet
receptacle strip.
Do not place containers holding liquid on or near the chamber door. If they spill, liquid may get into
the centrifuge and damage electrical or mechanical components.
Safety Against Risk of Fire
Certain electrical circuits within this centrifuge are protected by fuses against overcurrent condi-
tions. For continued protection against the risk of fire, replace only with the same type and rating
specified.
This centrifuge is not designed for use with materials capable of developing flammable or explo-
sive vapors. Do not centrifuge such materials (such as chloroform or ethyl alcohol) in this centri-
fuge nor handle or store them within the 30-cm (1-ft) area surrounding the centrifuge.