Spare channels for auxiliary data

Auxiliary analog input capability allows the user to acquire, record, display and gate
from user generated analog waveforms. The capability exists for two analog channels.
One channel sampled at the rate for respiration and a second channel sampled at the
rate for blood pressure. IBP sampling is 900 Hz. IBP sampling is twice the sampling
rate for respiration.
The auxiliary channel sampled at the respiration rate is labeled "USER RESP". The
channel sampled at the IBP rate uses the IBP (IBP3) connection on the IBP Module.
Parameters to adjust waveform gain, scaling, offset, polarity and labeling are set in the
MONITOR SETUP SCREEN (refer to Chapter 3) or by clicking on the associated
waveform display (refer to Chapter 9).
The connector pin outs for user RESP and USER IBP are listed below. (pin 1 is the
right most pin when viewing the connector):
Pin 1 n/c
Pin 2 +5 VDC
Pin 3 Signal +
Pin 4 Signal –
Pin 5 ground
Pin 6 n/c
Invasive Blood Pressure
Revision C, August 2013