System Setup
in milliseconds for the cardiac or respiratory cycle. Refer to Chapter 3 for additional
information regarding the main monitor display

Recommendations for daily operation

Connect the ERT Module Battery Pack to the charger at the end of the day. The battery
pack can be left on the charger when not in use. It can not be over charged.
The ERT Control/Gating Module can be powered on all the time.
The IBP Module - a) if you have power in the magnet room, use the 12 V power supply
provided. Make sure the power supply is not in a really strong fringe field. It will work in
a strong field, but will have a shortened lifetime. The module has an internal battery.
You can leave the battery switch and power switch in the on position all the time. b) if
you do not have power in the magnet room, then you need to operate the module on
internal battery and charge the battery overnight. Refer to Chapter 10.
The power to the Heater and Fan Modules should be turned off when not in use.
FO Temperature, FO Pressure and Pulse Oximeter Modules - We recommend
unplugging the 12 V power at the end of the day. Turning off power turns off the light
source for the temperature, pressure and oximetry probes.
Capnograph – turn off the main power switch when not in use.
Ventilator – turn of the main power and compressed gas supply when not in use.
Water Bed Heater System – turn off the power to the Heater Module and Circulation
Module when not in use.