User Ibp Setup Window - SA Instruments 1030 Operation Manual

Mr-compatible small animal monitoring and gating system
Table of Contents

Advertisement

Advanced Features
The waveform and scale labels are set in real time. They appear on the monitor display
as they are being entered in the USER RESP window.
The analog to digital converter (ADC) has 1024 bits. The user can select the scaling
and offset between the ADC scale and the displayed scale. The displayed max and min
scaled values can be changed by right and left clicking the values on the monitor
display. The step size for this adjustment is set in the Range Adjust Step Size box.
Application of the high pass filter can be useful to remove a DC offset. Selecting
"Output to Resp LED" flashes the USER RESP Module's LED for each User Resp gate.
Analog Input should be selected for the signal Source. The other option T-Resp™
which generates a respiration signal from a temperature sensor is under development
and has not been released.
Three gain stages allow overall signal gain to be set between 80 and nearly 20,000 with
the option to change signal polarity. The middle gain stage can also be set to AutoGain
to automatically track a signal of varying amplitude.
The User Resp software can interpret the waveform in the following ways:
• As a respiration waveform displaying respiration rate and period
• As a slowly changing analog value, like temperature, displaying the value of the
measurement
• As a capnograph waveform displaying the respiration rate and end-tidal CO2.
To make the selection for how to interpret and display the information, execute PC-sam
from the desktop and select the option to display the "full setup screen", then make the
selection at the bottom of the USER RESP SETUP window "Rate/Prd". "value" or
"Rate/EtCO2". The corresponding measurements will be displayed to the right of the
USER RESP waveform. The data will also be recorded in the Trend file.
User Resp gates are automatically generated for the User Resp waveform.
presence of gates generates values for User Resp respiration rate and period. The
gates can be displayed on the respiratory sweep display and included in the algorithm
for generation of output gates.

USER IBP SETUP window

Right click on the User IBP waveform to open the USER IBP SETUP window. The IBP
connector on the IBP Module can be used as an IBP channel (IBP3) or as a user
auxiliary analog input channel User IBP.
15-4
Revision C, August 2013
The